Last updated on 23-8-2019 by Anonymous (non vérifié)
Peer reviewed scientific article
SCIENSANO
Mots-clés
Résumé:
The ultrasonic pachometer has been introduced recently into our laboratory for use in the eye-irritation test using enucleated rabbit eyes. We have compared the performance of this kind of pachometer with that of the optical pachometer. The results for corneal swelling, reported by Burton et al. (1981) and obtained using an optic pachometer, were compared with those obtained using an ultrasonic pachometer to test the same 11 substances. The mean corneal swelling 240 min after application of the test substances to four eyes resulted in a very good correlation (r = 0.984) between the optic an…