Development and validation of an ultra-high performance liquid chromatography-tandem mass spectrometry method to measure creatinine in human urine.

Last updated on 7-4-2021 by Jill Alexandre

Public Access

Published

Peer reviewed scientific article

Engels

Samenvatting:

Despite decades of creatinine measurement in biological fluids using a large variety of analytical methods, an accurate determination of this compound remains challenging. Especially with the novel trend to assess biomarkers on large sample sets preserved in biobanks, a simple and fast method that could cope with both a high sample throughput and a low volume of sample is still of interest. In answer to these challenges, a fast and accurate ultra-high performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method was developed to measure creatinine in small volumes of human…