Last updated on 23-8-2019 by Anonymous (non vérifié)
Peer reviewed scientific article
SCIENSANO
Mots-clés
Résumé:
Although licensed batches of an enzyme-linked immunosorbent assay (ELISA) for Aujeszky’s disease virus (ADV) were used, and the assays were performed within an ISO/IEC 17025 accredited quality control system, certain routine runs of the ADV ELISA were not validated using the quality system criteria, even when all technical parameters were controlled. Incubation at different temperatures and batch composition were identified as parameters that could result in non-validated assays/runs. Therefore, the effect of incubation temperature and batch composition on the analytical sensitivity of the …