Performance of automated measurement of antibodies to cyclic citrullinated peptide in the routine clinical laboratory97

Last updated on 23-8-2019 by Anonymous (non vérifié)

Public Access

Published

Peer reviewed scientific article

Résumé:

OBJECTIVE: To evaluate the performing technical and clinical characteristics of an automated system for routine measurement of anticyclic citrullinated peptide antibodies (aCCP), a new marker for rheumatoid arthritis (RA). MATERIAL AND METHODS: Reproducibility, repeatability and linearity of aCCP, as measured by an automated fluorescent enzyme immunoassay (FEIA/Phadia), were evaluated and compared with the performance of a manual ELISA technique (Axis Shield Diagnostics). Clinical verification of both methods included estimation of sensitivity in RA patients (n = 42) and specificity in well…